16 Technical specifications

01 Dimensions: 62.0 mm (Width)
× 37.0 mm (Depth) × 32.0 mm
(Length) Weight: 42.5 g (includ-
ing two AAA dry-cell batteries)
02 Peak wavelength range of light
emitted by the probe: red light 660
nm ±3; 905 nm ±5 infrared light.
03 Maximum optical output pow-
er of the probe: 1.2 mw for
infrared light (905 nm).
04 Date of manufacture: see label
05 Normal working condition
Operating
temperature
Relative humidity
Atmospheric
pressure
Voltage rating
M-PO100-001.indd 9
M-PO100-001.indd 9
06 Default values and warn-
ing conditions
SETTING
Oxygen sat-
uration
Heart rate
Warning
condition
5°C to 40°C
07 Technical settings
(41°F to 104°F)
15% to 80%,
SETTING
no condensation
saturation
70 kPa to 106 kPa
DC 3.0 V
saturation
VALUE
Upper limit: 100
Lower limit: 94
Upper limit: 130
Lower limit: 50
When the warning switch
is activated and the ac-
tual measured value
exceeds the predefined
range of warning set-
tings, the oximeter emits
a warning sound.
VALUE
Oxygen
35% to 100%
Heart rate 25 bpm to 250 bpm
Oxygen
1%
Heart rate 1 bpm
SETTING
VALUE
±2% (70% to 100%)
Oxygen
No require-
saturation
ment (≤ 69%)
Heart rate ±2 bpm
Upper limit:
Oxygen
50% to 100%
saturation
Lower limit:
50% to 100%
Upper limit:
25 bpm to 250 bpm
Heart rate
Lower limit:
25 bpm to 250 bpm
Oxygen
±1% of the pre-
saturation
defined value
The highest of ±10%
Heart rate
of the predefined
value and of ±5 bpm
PI
Low PI
Min. 0.3%

17 Safety category

• Type of protection against electric
shocks: internally powered equipment.
• Level of protection against electric
shocks:
type BF applied part.