How to adjust the basic parameters of the movement?

Beginning
To stop the unit
To display the extension or flexion limit of the
movement
To change the limit if necessary
To validate the new value, press another key
or wait more than 3 seconds.
Or
Or to display pause in extension or flexion limit
To change the pause if necessary
To validate the new value press another key
or wait more than 3 seconds. The display
shows the selected mode

How to use the WARM UP key?

Warm Up rules.
The Kinetec Spectra starts at 70% of the full ROM, increasing 5% of the range every other
cycle until the pre-set ROM is reached.
Beginning
To stop the unit
To select the Warm Up mode
Start the movement
Comments:
- The parameters modification should always be done while the unit is sopped.
- Pauses and By-Pass mode are not available during warm up cycles.
- The Warm Up cycles are only working when the movement is firstly run.
- To start new Warm Up sessions press twice the « Warm Up » key.
- The calculation mode used enables reaching the pre-set ROM in approximately seven full cycles.
Example:
For a Warm Up treatment with a pre-set ROM from 0° to 100°.
The first cycle starts from 15° to 85° to 15° and increase 5% each cycle.
Spectra & Spectra DC
Keys to press
30 STOP
30
Or
40 FLEX
40 FLEX
Or
LOAD _ _ _ _ _ _
TIMER 00H00MIN
PAUSE FLEX
PAUSE EXT
PAUSE EXT
Or
40 STOP
Keys to press
40 STOP
WARM UP
Please Wait.P
Nom
Prenom
WARM UP
40 STOP
WARM UP
40 RUN
EN
Display
KINETEC
Check if the hand control is not
45 70
blocked (See page 5).
KINETEC
EXT
45 70
The value blinks.
KINETEC
45
70
KINETEC
The new value blinks.
45
110
While the value blinks press
KINETEC
SPEED 2
the
KINETEC
to change if necessary.
KINETEC
KINETEC
0S
Successive presses on this key selects
the pause at the extension or flexion
KINETEC
limit.
0S
KINETEC
The new pause value blinks.
15S
KINETEC
The unit is ready to start with the new
45 110
parameters.
Display
KINETEC
Check if the hand control is not
45 110
blocked (See page 5).
xx
Spectra DC only.
xx
To change the movement value if
45 110
necessary (see page 6).
The value change at the speed of
50
110
the movement.
Remarks
or
key
100°
50°
Remarks
7/16