Chauvin Arnoux C.A 41 Notice De Fonctionnement page 39

Mesureur de champ électrique large bande
Masquer les pouces Voir aussi pour C.A 41:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Variations in the operating range
Distortion magnitude
Ambiente temperature
Humidity
Power supply
Frequency of field
Level of field
Interchangeability of the
sensors
All distortion magnitudes
Limit of range
0 to 50°C
10 to 90 %
no condensation
7.5 to 11 V
20 to 500MHz
1 MHz to 1 GHz
1 MHz to 2.5 GHz
0.1 to 10 V/m
0.1 to 100 V/m
0.1 to 200 V/m
20 to 500MHz
1 MHz to 1 GHz
1 MHz to 2.5 GHz
Level from
0.1 to 200 V/m
from 0 to 50 °C
from 10 to 90 % HR
power supply
from7.5 V to 9 V
frequency from
1 MHz to 2.5 GHz
Level from
0.1 to 200 V/m
39
Magnitude
distorted
all magnitudes
all magnitudes
all magnitudes
all magnitudes
all magnitudes
all magnitudes
all magnitudes
analog output
from 0 to 20 V/m
Alarm detection
threshold
MAX variation
0.3 % / °C
of the reading
± 0.5 V/m per 10 °C
< 0.1 V/m
0.05 %/V
± 1 dB
± 1.5 dB
± 2 dB
± 0.5 V/m
± 1 dB
± 2 dB
± 1 dB
± 0.5 dB
± 1 dB
of the display value
± 0.2 V/m
of the displayed
value

Publicité

Table des Matières
loading