Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The BC 31 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BC 31 should assure that it is used in such an environment.
IMMUNITY
IEC 60601
test
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection
from structures, objects and people.
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer to any
part of the BC 31, including cables, than the recommended separation distance cal-
culated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Not
applicable
d = 1.2
80 MHz to 800 MHz
d = 2.3
3 V/m
Where P is the maximum output power rating of the transmitter in watts (W) accor-
ding to the transmitter manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fi xed RF transmitters, as determined by an electromagnetic site
a
survey,
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
800 MHz to 2.5 GHz