DANGERS, WARNINGS, CAUTIONS
^ Danger
Failure to observe the following procedures may result in serious personal injury
• Not for use in explosive or hazardous locations. This product is not intrinsically safe.
• Contains built-in lithium polymer battery. Do not incinerate or dispose of in fire. Do not
disassemble, alter or re-construct.
^ Warning
Failure to observe the following procedures could damage the instrument.
• Read the manual before operation.
• Do not store in temperatures outside -20°C to 60°C (-13°F to 140°F).
• Do not immerse in liquids.
• Condensation may damage your instrument.
• Substitution of components may impair the accuracy of the instrument. Repair should be
done by authorized service personnel only.
^ CAUTION
• The battery in this instrument has limited shelf-life, even if never used.
• A non-condensing environment is required for proper instruments.
• Do not charge battery outside the range of 0°C to 40°C (32°F to 104°F).
Battery run-time may be reduced when operating at temperatures lower than 0°C (32°F).
• It is recommended that the SE-400 Series is charged for approximately 4 hours before first
use.
• The SE-400 Series Sound Level Meter may not be sufficient to carry out a full risk
assessment of workers exposed to hazardous noise levels.
Figure 1.
A. Softkeys: Four softkeys used as additional menu options with specific screens.
B. Backlight Key: Used to illuminate the background of the display screen.
C. Up Arrow Key: Used to toggle through menu/setup options and scrolls through the values
on the measurement screen.
D. Left Arrow Key: Used to navigate through menus and setup parameters.
E. Run/Stop Key: Used to run and stop a logged session.
F. Right Arrow Key: Used to navigate through menus and setup parameters.
G. Enter Key: Used to select menus and confirm setup parameter changes.
H. Down Arrow Key: Used to toggle through menu/setup options and scrolls through the
values on the measurement screen.
I. On/Off/Esc Key: Used to power on, power off and/or escape or move back one screen.
J. Windscreen: Used to reduce noise effects from wind with minimal effects on the readings.

INTENDED USE

The SE-400 Series Sound Level Meter is intended to measure sound pressure levels in air
and provide sound and noise detection measurements. It is fitted with12 keys for powering
on/off, viewing and/or running/stopping measurements, navigating and setting parameters
(See Figure 1 for more information). For suitability and proper use follow local regulations and
refer to all information supplied. For more information contact a safety professional/3M
representative (Local Contact Details).

Getting Started

Attaching the Preamp
The SE-400 Series, modelSE-401 is equipped with a removable preamp and is shipped
unattached to the SLM.
1. Place the preamp connector over the mating connector at the top of the instrument. Gently
press down while rotating the preamp ring until the preamp connector drops into place.
2. Rotate the preamp ring until secure.
Charging
TheSE-400 Series is charged using the supplied USB charger or personal computer's USB
port. A fully depleted battery will require approximately8 hours to fully charge when using the
external battery charger.
Status of battery icon: When charging, the battery indicator will scroll, shading in the
indicator. When the battery is fully charged, the battery indicator will stop scrolling.
Powering On/Off
1. To turn on the SE-400 Series, press the
2. To turn off the SE-400 Series, ensure the meter is stopped. Press and hold the
until the measurement screen disappears.
Calibrating
It is advisable to perform a field calibration before taking measurements to ensure your
measurements are accurate in the current environment.
1. Ensure theSE-400 Series is turned On, stopped, and not logging.
2. Attach the calibrator and calibrator adapter to the SE-400 Series. Set the calibrator to1000
Hz and114 dB (if it is selectable).
3. Turn On the calibrator.
4. From the measurement screen, press the corresponding softkey from the
5. Press the CAL softkey or select Calibration (See Figure 2-A.)
6. In the calibration screen, press CAL softkey or press
7. Press the corresponding
softkeys to adjust the measurement to114.0 dB. (See
Figure 2-C.)
8. Press the
key to save the changes or escape to abort.
9. Press the
key twice to return to the main measurement screen.
key until the start screen appears.
t
menu.
key. (See Figure 2-B.)
Figure 2-A.
Windscreen
The SE-400 Series features a removable windscreen. This is primarily designed to reduce noise
effects from wind with minimal effect on the readings. It is recommended to use the instrument with
the windscreen on at all times in order to protect the microphone.
Positioning
For area monitoring and general sound level sessions, position the instrument in the field at a0°
angle towards the noise source.
Settings
1. The frequency weighting and the time response may be changed on the instrument by pressing
the corresponding FS or ACZ softkeys when in stop mode from the measurement screen. (See
Figure 4-B.)
2. To access the settings menu, select the
• To change the log interval, exchange rate, peak weighting, language, and/or contrast, select
time/date, language settings by pressing
(See Figures 3-A and 3-B.)
Figure 3-A.

Operating

Run, stop, and view results
1. To start a session, press the
the current duration of the session.
• The multi-measurement screen (see Figure 4-A) will appear with the SPL displayed.
2. Press the
3. Press the
key to view single measurement screen(see 4-B).
Press the
Figure 4-A.
Viewing measurements in DMS
1. Optional: To view the logged data in charts, tables, and customized reports, download the data
into 3M™ Detection Management Software (DMS).
Figure 5-A.
key
IMPORTANT: For appropriate use of hearing protection devices, consult your safety professional,
applicable local standards or contact 3M (Local Contact Details). Note: Hearing protection sold
separately. This product complies with the requirements of IEC61672-1 with the exception of
clause 9.1c - "An instruction manual shall be provided as a printed document in one or more parts".
A printed copy of the instruction manual is available on request.
3
Figure 2-B.
Figure 2-C.
setup icon with the corresponding softkey.
g
keys and press
Figure 3-B.
key. The run icon will appear with a run-time clock indicating
i
arrows to scroll through measurements.
arrows to scroll through measurements.
Figure 4-B.
LASmx value
Time Response
Frequency Weighting
key.