2. The last measurement reading will automatically be
shown on the display for 2 seconds with the «M» icon 7.
3. When the «°C» or «°F» icon is flashing, a beep sound is
heard and the thermometer is ready for measurement 8.
4. Straighten the ear canal by pulling the ear up and back
to give a clear view of the eardrum.
 For children under 1 year; pull the ear straight back.
 For children of 1 year to adult; pull the ear up and
back.
Also refer to the short instructions at the front.
5. While gently pulling the ear, insert the probe snugly into
the ear canal (max. 1 sec.) and immediately press the
START button 2. Release the button and wait for the
beep sound. This is the indication that confirms the end
of measurement.
6. Remove the thermometer from the ear canal. The
display shows the measured temperature 9.
NOTE:
 To ensure accurate readings, wait at least 30 sec.
after 3-5 continuous measurements.
 Accumulation of ear wax on the probe can result in less
accurate temperature readings or cross infection
between users. Therefore, it is essential that the
probe is clean before each measurement. For
IR 100
cleaning, follow the instructions in the «Cleaning and
Disinfecting» section.
 After cleaning the measuring sensor 1 with
alcohol, wait 5 minutes before taking the next
measurement, in order to allow the thermometer to
reach its operating reference temperature.
 10 short beeps will sound when the temperature is
higher than 37.5 °C (99.5 °F) in order to alert the patient
that he/she may have fever.
 For an infant, it is best to have the child lying flat with his/
her head sideways so the ear is facing upwards. For an
older child or adult, it is best to stand behind and slightly
to the side of the patient.
 Always take the temperature in the same ear, since the
temperature readings may be different from ear to ear.
 Wait for a few minutes to take the ear temperature after
sleeping.
 In the following situations it is recommended that three
temperatures in the same ear be taken and the highest
one taken as the reading:
1. New born infants in the first 100 days.
2. Children under three years of age with a compromised
immune system and for whom the presence or
absence of fever is critical.
3. When the user is learning how to use the thermometer
for the first time until he/she has familiarized himself/
herself with the instrument and obtains consistent
readings.
4. If the measurement is surprisingly low.

6. Changing between Celsius and Fahrenheit

This thermometer can display temperature measurements in
either Fahrenheit or Celsius. To switch the display between
°C and °F, simply turn OFF the unit, press and hold the
START button 2 for 5 seconds; after 5 seconds, the current
measurement scale («°C» or «°F» icon) will be flashing on
the display AL. Change the measurement scale between °C
and °F by pressing the START button 2. When the measure-
ment scale has been chosen, wait for 5 seconds and the unit
will automatically enter the «ready for measuring» mode.

7. How to recall 12 readings in Memory Mode

This thermometer can recall the last 12 readings.
 Recall mode AM: Press the START button 2 to enter
Recall mode when power is off. The memory icon «M»
will flash.