LaserLiner MultiFinder Pro Mode D'emploi page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
Tip 1: The position between the two markings is the mid-point
of the metal object. Through the high measuring sensitivity,
thick metal objects appear broader than they are in real life.
Therefore move the device over the newly found object as
shown in Image B. The device is calibrated automatically during
this process. Manual calibration should be performed near
the place found last as shown in Image C. Repeat this step as
needed.
Tip 2: The position where you start is important: First place
the device in a position where you know there is no metal.
Otherwise, the message "ERROR" will appear in the display.
To remedy: Move the device to another position a few
centimetres away and start measuring again.
Tip 3: In the case of complicated applications, e.g. ribbed
steel, scan the surface both horizontally and vertically.
Tip 4: Flexible floor and wall heating pipes which contain
a metal foil and are located near the surface may also be
detected. Test for this function in places where you know
the position of such pipes.
Note: If an object is deep inside a wall, the device may not
indicate it clearly.
7
Measuring in STUD-SCAN mode
Detecting wooden wall beams and joists as well as metal in
drywall structures, e.g. under gypsum fibreboard, wood panels
or other non-metallic cladding.
– Select STUD-SCAN (button 4).
– Now follow the instructions on the VTN display.
– ON WALL: Place the tool against the wall.
– PRESS CAL: Press the calibration button (5) and wait until
calibration is completed: CAL OK
– MOVE: Move the tool slowly across the surface.
Manual
calibration
1.
8.
MultiFinder Pro
2.
6.
3.
7.
4.
5.
EN
15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Multifinder plus

Table des Matières