Quick start guide
English
Do not lift or handle the instrument by the black external
Temperature, %RH, pressure sensor housing!
1
Press
to turn on the instrument then
2
Si-AQ Comfort & Si-AQ Pro: at initial startup, allow the CO
to warm up and stabilize.
Be sure to stand arms length from the instrument while taking measurement to
ensure that you are not affecting readings.
Si-AQ Comfort & Si-AQ Pro: if applicable, attach the optional probe and sampling
3
line to the monitor.
4
Press
to turn Pump On/Off. When Pump is OFF, no sample is being made.
5
Si-AQ Comfort & Si-AQ Pro: zeroing sensors using outdoor fresh air (except CO
The user should be at least 20-30 feet (6-10 meters) away from any buildings, parking lots, highways, and
any other possible exhaust sources. If in doubt of the ambient air, please use the ZERO AIR cylinder.
• Press
. When you turn the instrument on in clean ambient air, wait at least one (1)
minute to allow the monitor to warm up before carrying out the zero calibration. Zero Gas
Sensors: this will set the zero point of gases based on the clean ambient conditions.
• Have the cursor (reverse color) point to Zero Gas Sensors. Press OK to
execute a zero cycle of all the gas sensors. This procedure should last 30 seconds.
• At the end of the zero period the instrument reads the output of all gas sensors (with
the exception of the CO
that is set to 20.9%.
• If no error messages appear at the end of the countdown, proceed with your measurements.
Si-AQ VOC: zeroing VOC & Gas Sensors using Si-AQ VOC Zero Filter (except CO
in ambient air conditions
Let the instrument run in ambient conditions for minimum of 5 minutes BEFORE Connecting the Si-AQ VOC
Zero Filter and performing the ZERO.
• Press
• To zero the gas sensors (except CO
the Si-AQ VOC Zero Filter filter into the Middle Sample In
"A" inlet at the bottom of the Si-AQ VOC and ensure that
the opening on the bottom end of the filter is not blocked.
Allow the unit to run with the filter connected for at least 30
seconds before continuing, this ensures that any potential
contaminants inside the unit are purged.
• Zero Gas Sensors: this will set the zero point of VOCs and up to two (2) other gases
from the following list: CO, NO, CH
the Si-AQ VOC Zero Filter.
to display measurements.
sensor up to 1 minute
2
sensor) and sets them all to zero, except the Oxygen (O
2
) in Ambient Air, insert
2
O and H
S values based on the cleaned air using
2
2
With Filter Connected, press OK to execute a Zero cycle of all the gas sensors while
the instrument is running in ambient air. This Zero cycle lasts 30 seconds.
• Press
• At the end of the zero period, the instrument reads the output of all gas sensors
(except CO
to 20.9%. If no error messages appear at the end of the countdown, place the
Si-AQ VOC Zero Filter back into the included sealable bag to prevent moisture
from entering it.
CO
should ALWAYS be ZERO Calibrated with a Cylinder of ZERO AIR. Do not use
2
ambient air for CO
6
Si-AQ VOC: sensor protection with AQ Moisture Filter
All IAQ devices equipped with PID or optical sensors such as VOCs and/or CO
be protected from potential moisture, humidity and/or particulates in the air. Each Si-AQ VOC
unit is equipped with one (1) disposable AQ Moisture Filter (Si-AQ Moisture Filters).
• After Zeroing of the Si-AQ VOC unit using the Si-AQ VOC Zero
)
2
Filter is complete, insert the provided AQ Moisture Filter into
the Middle "A" inlet at the bottom of the Si-AQ VOC and ensure
that the opening on the bottom end of the filter is not blocked.
• Leave the AQ Moisture filter connected during your analysis/
sampling for better protection of the sensor.
Storing Data to Internal Memory (See Chapter 7 in the complete user manual)
7
• Press
• Select Store Current Buffer to store a single complete set of data or one of the
)
other Buffer Storage options depending on your needs. To store data in a different location,
2
use the UP, DOWN, & OK keys to select a new tag. Empty tags show the word empty
• Select Review Buffer to view previously stored data
• Select Name Buffer and Erase Buffer as needed.
)
2
8
Start periodic data storage: this will turn on the periodic (long-term) store function. In
this mode, the unit will continuously store data at an interval displayed on the next line.
• In the Data Storage menu, highlight Select Interval. The time between
each store is set here. This can range from 10 seconds to 60 minutes.
• To start data storage: highlight and press OK to Start Periodic Store
• To stop data storage: highlight and press OK to Stop Periodic Store
9
Reattaching the Temperature/%RH probe
• If the probe becomes detached, reinsert into the round inlet
at base of the unit.
• Si-AQ VOC: when connecting the probe, be careful to align
the probe so that the black stripe is oriented to the top/front
of the instrument.
and have the cursor (reverse color) point to Zero Gas Sensors
sensor) and sets them all to zero, except Oxygen (O
2
ZERO sensor calibration (See Chapter 9 in the complete user manual)
2
) that is set
2
sensors should
2