Installation Instructions; Installation Configurations - SciCan HYDRIM C51w Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4. Installation Instructions

To open the wash chamber door, pull the ring
inside the chemical door.
Tools and supplies required for installation:
• Slot screwdriver
• Channellocks
Ensure that HIP Cleaning Solution (instrument wash chemical) is
available. All other supplies are included with the HYDRIM unit.
Specifications:
North America Voltage and
Frequency: 208-240V, 60 Hz
International Voltage and
Frequency: 220-240 V, 50 Hz
Rated load: 2 kW
Circuit breaker:
10 A per phase
Utility Hook-ups
Hose /
Length /
Cord
Diameter
Hot inlet
1.9 m / 6.2 ft. 1.5 m / 5 ft.
3/4"
Cold Inlet
1.9 m / 6.2 ft. 1.5 m / 5 ft.
3/4"
Drain
1.5 m / 5 ft.
3/4"
Electrical
1.8 m / 6 ft.
AWG 18-3
* unit will function with water pressure down to 0.5 bar/ 7 psi.
drain hose
cold water
inlet hose
hot water
inlet hose
Page 4
Height:
475 mm/18.75"
Width:
605 mm/23.8"
Depth:
460 mm/18.25"
Depth with
door open: 780 mm/30.75"
Weight:
34 kg/75 lbs
Running
Noise:
60 dBA
Max. Distance
Water
Shut
Distance from
Pressure
inlet / drain
(optimal)*
valve
1-10 bar
14.5 - 145 psi
1-10 bar
14.5 - 145 psi
inlet valves
The electrical outlet
should not be located
directly behind the unit
and should be accessi-
ble. The powercord must
be routed away from the
back panel and hotwater
inlet hose.

Installation Configurations:

If the HYDRIM is installed in a Sterilization Center,
the manufacturer of the Sterilization Center should
allow a 75 mm/3" space at the top, back and both
sides of the HYDRIM. This will facilitate installation,
leveling, and service access to the HYDRIM.
The cabinetry will need to be vented and allow
the air exhaust from the dryer to escape. At least
75mm/3" of space should be provided behind the
rear panel.
Do not move the HYDRIM into place by maneuver-
ing the open wash chamber door. This may cause
to door to become misaligned and leak.
Option 1. Cabinet / Sterilization Center
-off
Yes
Option 2. Sliding Shelf
Yes
Option 3. Counter Top

Publicité

Chapitres

Table des Matières
loading

Table des Matières