Calibration checks - Testing instructions
(for trained specialist staff only)
Testing
1.) Remove batteries.
2.) Hold start/stop button
down and insert batteries.
3.) Release start/stop button
The monitor is now ready for testing;
the current pressure is displayed in the
SYS and DIA fields.
4.) Perform test for deviation of pressure
display and pressure circuit integrity
(observe setting time for cuff – at least
30 seconds) in the usual manner.
5.) Switch off monitor by pressing the
start/stop button
establish all original connections.
pressed
1
C) Safety seal
.
As a safeguard, the upper and lower
1
parts of the housing can be joined
with a safety seal. Alternatively, a hole
drilled in the bottom of the monitor for
the connecting screw for the housing
can be sealed for safety.
or maintenance while it is in use.
1
The device must not undergo service
again and re-
EN
79