This chapter includes setting of alarm values.
The alarm system is divided into 2 systems. The one system only triggers local alarms, which
means that the error codes are only shown at the display.
The other alarm system triggers both alarms at the display, but it also activates the remote
alarm outlet. Each alarm system operates independently of each other.
If a change of the setting is required, get access to the menu by pushing
than 3 seconds. The values are changed by pushing the
saved by pushing
Menu access P + 1
Local alarm setting
External alarm setting
Temperature offset
(sensor calibration)
Escorting alarm limits
No. of defrosts
765041528
Gram-webshop.nl
P
. To exit the menu, push
LAL
LHL
LHd
dA
dAd
BU
EAL
EHL
ELL
EHd
ELd
CAL CA
ALL
FAS/ESC
dEF
4
Rev. 000
-
+
or
.
+ / -
°C
Setting the upper alarm limit. At
alarm, the display shows: A2
min.
Time delay for the upper alarm limit.
On/off
Activation of local door alarm. At
alarm, the display shows: A1
(1=on / 0=off)
min.
Time delay for the door alarm.
On/off
Activation of buzzer. The buzzer
sounds at alarms A1, A2. (1=on /
0=off)
°C
Setting the upper alarm limit. At
alarm, the display shows: A3
°C
Setting the lower alarm limit. At
alarm, the display shows: A4
min.
Time delay for the upper alarm limit.
min.
Time delay for lower alarm.
-5...+5 K
K
Activation of escorting alarm limits.
FAS= fixed limits / ESC = limits
following the setpoint
Number of defrosts in 24 hours.
>3 sek.
P + 1
for more
keys. The new setting is