1.3 Standard settings / Factory settings
• Note the following default settings of the data logger before first use. By using the
DE-LOG-Graph software, the setting parameter can easily be changed:
1. Description: blank (max. 16 characters)
2. LCD-Snooze mode:
3. LCD-Snooze after Sec.: 30
4. Alarm settings for temperature: -30,0°C 60,0°C
Alarm settings humidity: 0,0% 100,0% (Log20 only)
5. Alarm delay:
6. Alarm cumulation: off
7. Alarm reset:
8. Temperature unit: °C
9. Waiting for manual start:
10. S ingle use only:
11. Measuring interval: 15 Minutes
12. C ycle memory: (if the memory is full the oldest measurement will be overwritten)
= Default
1.4. Marking
• Log10/20: CE-conformity
• Log10 only: EN 12830, EN 13485, Suitability for storage (S) and transportation (T)
for food storage and distribution (C), Accuracy classification 1 (-30..+70°C),
according to EN 13486 we recommend a recalibration once per year.
12
2. Operation
• For configuring the data logger, please install the Software DE-LOG-Graph on a PC.
2.1 USB-Port
• When the Software Installation has been completed please connect the PC with
data logger via USB-cable. For detailed information please read the manual of the
DE-LOG-Graph-Software.
2.2 Panel and display (Fig. 1)
Log10/20 has a large display, one LED and one button.
A: LCD-display indicates temperature, humidity (Log20 only), Low bat-warning,
Max-Min-Avg-measurements, status information
B: Start-Stop-button
C: LED: red
D: USB-port (with rubber cap)
2.3 Handling the Start-Stop-button
• Depending on the setup configuration, you can start or stop the data logger via the
Start-Stop-button. You have to press and hold the button for 3 seconds. When it
starts, the display indication will switch from STOP to LOG.
• MAX-, MIN-, Average values:
At the bottom line of the display, you will see the Average (AVG)-, Minimum (MIN)-
and Maximum (MAX) temperature of the recorded measurements. If the data logger
is not started it will display --- instead of AVG, MIN or MAX temperature.
13