7.12
Torch light
•
Press the
will turn itself off
•
Turning on of torch light will change operating mode on
SEFRAM 66
7.13
Frequency test (SEFRAM 66)
•
Switch into frequency measurement by short pressing 4 times
the
button or repeat short pressing until Hz symbol is
shown on LCD. Fr and Hz symbol should be on LCD. Connect
both test probes to the AC voltage under test. Frequency from
1 Hz to 800 Hz can be shown on LCD
Frequency measurement is possible for voltages >10 VAC
•
The level of voltage will be shown only on the bar graph
7.14
Self test
•
LEDs up to 50 V (SEFRAM 62), all LCD segments and vibration
motor (SEFRAM 66), torch light, ELV indication and buzzer are
turned on for a second after battery replacement or selftest
activation
Selftest can be activated:
Shorting L1(-) and L2(+) probe tips while device is turned
OFF – leave device OFF for 30 seconds before shortening tips
7.15
NCV Test (SEFRAM 66)
•
Switch into NCV measurement by short pressing 5 times the
button. LCD will show NCV symbol
•
The NCV function is used to verify that the wire is energized
•
Hold the voltage tester with the sensor against the wire or
cable. The voltage tester indicates the strength of the signal
digitally on the LCD screen
7.16
HOLD Function (SEFRAM 66)
•
Long press (2 seconds) on
and freezes display value. With short press on 'torchlight/
function' button freezing of display is released. When HOLD
function is active
7.17
Backlight (SEFRAM 66)
•
White Backlight is on all the time while tester is turned on
•
Backlight will be green when voltage <50 VAC or <120 VDC
•
Backlight will be red when voltages >50 VAC or >120 VDC and
for single pole detection
7.18
Battery Replacement
Remove the probes from any testing point, when opening the
battery case. Batteries are dead when the continuity test with
both test probes connected cannot be done anymore.
Follow the procedure below and replace batteries with new ones
(type IEC LR03 1.5 V).
•
Unscrew the battery door
•
Pull out the battery door and replace the batteries
•
Insert new batteries according to the printing on the battery
door
•
Re-assemble battery door.
•
Confirm that the battery door case is properly locked prior to
measurements
8
Cleaning and storage
Tester does not need any special maintenance if used according
to user manual.
Remove tester from all test points before cleaning.
Use a lightly damp cloth with neutral detergent for cleaning the
instrument. Do not use abrasives or solvents.
Do not expose the instrument to direct sun light, high tempera-
ture and humidity or dewfall.
Remove batteries when the instrument will not be in use for a
long period.
button to turn on the light and after approx. 30 s it
symbol will be shown on LCD
www.sefram.fr
SEFRAM 62 - 66
button activates HOLD function
19