To clean
- Open and empty the cleaning kit basin
(optional Cat. No. 081500).
- Place the trainer head face down in the cleaning
basin.
- Connect the two lung tube connectors to
the double Y-piece in the tube system.
- Connect the oesophagus to the single connection
in the tube system.
- Insert the cleaning syringe into the valve opening
and place the free end of the tube system into
the basin.
- Fill the basin with liquid to just underneath
the edge. Circulate liquid through the manikin
airways by pumping the syringe plunger.
- Use this setup to go through the following steps:
1. Use a Disinfectant solution*. Airways should
remain completely filled for at least 10 minutes.
2. Use fresh water to remove disinfectant solution.
After each step lift the head clear of the liquid
to allow drainage.
3. Allow to dry completely before reconnecting head
to body part.
* Virkon can be used.
Depending on local procedures you may prefer to use a
sodium hypochlorite solution freshly prepared for single
class use.This solution should have at least 500 ppm
free available chlorine, i.e. 22 ml of 2.5 - 5.0 % house-
hold bleach per 500 ml of water. Local cleaning and
handling procedures must then be followed.
PARTS LIST
250-20150
Body part
250-20250
Lungs/stomach (pkg. 3)
250-20350
Base plate
250-20450
Cover
250-21050
Airway Lubricant (45 ml)
081500
Cleaning kit
082022
Stomach valves (pkg.10)
083300
Head cpl.
083310
Face skin
083320
Head without face skin
252090
Airway lubricant (180 ml)
6