Table 2 – electromagnetic IMMUNITY – for ZAM-01,02,03
Guidance and manufacturer's declaration – electromagnetic immunity
The ZAM-01,02,03 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ZAM-01,02,03 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge(ESD) EN
61000-4-2
Electrical fast
transient/burst EN
61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
EN 61000-4-11
Power frequency
(50/60 Hz)
magnetic field EN
61000-4-8
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT) (
>95% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 s
3 A/m
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT) (
>95% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 s
3 A/m
46
Electromagnetic environment-
guidance
Floors should ne wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environmental
Mains power quality should be
that of a typical commercial or
hospital environmental
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the ZAM-01,02,03 requires
continued
operation
power mains interruptions, it is
recommended that the ZAM-
01,02,03 be powered from an
uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
environment.
during
hospital