Brandt Transferpette electronic Mode D'emploi page 63

Pipette à piston
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 82
g.
h.
1 x
i.
1 x
j.
1 x
k.
1 x
m.
Transferpette® electronic 997209
g.
Set the number of steps: Pressing the arrow key (+/-)
sets the number of steps.
ð
'Steps' continues to flash.
h.
Confirm number of steps: Press 'Enter' key.
ð
The display shows the set number of steps.
i.
Aspirate liquid: Press the pipetting key once to aspirate
the liquid.
j.
Dispense liquid: Each time the pipetting key is pressed,
a dispensing step is performed. The arrow in the display
points downwards (dispensing). The 'Step' display
shows the number of steps remaining.
k.
Trigger blow-out: After the last dispensing step, press
'Enter' key.
ð
The display now once again shows 'blo' for blow-out.
l.
End dispensing: By pressing the pipetting key once, the
blow-out is triggered, and the residual liquid is dis-
pensed.
m.
After dispensing the residual liquid (blow-out), the dis-
play jumps back to the set mode (start position).
5 Pipetting programs
63
Operating manual

Publicité

Table des Matières
loading

Table des Matières