Pipetting; Piezas De Repuesto - Brandt Transferpette electronic Mode D'emploi

Pipette à piston
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 82
b.
+
-
c.
1 x

4.4 Pipetting

The volume is factory-set to the respective nominal volume of the Transferpette® electronic and can
be easily and quickly changed individually, see Set volume, p. 51.
b.
1 x
Transferpette® electronic 997209
b.
Press the arrow keys (+\−) to select the speed level (e.g.
level 2).
ð
'Speed' continues to flash.
c.
Press 'Enter' key.
ð
The display returns to the basic state of the respective
set mode. Here, for example the display of the standard
PIP mode.
a.
Attach tip vertically:
Use the correct tips, in accordance with the volume
range or color code!
Make sure that the tips are firmly in place and leak tight.
When using the flexible pipette shaft, attach an alterna-
tive ejector adjustment clip if necessary.
Pipette tips are disposable products!
b.
Aspirate liquid: Hold the instrument vertically, and im-
merse the tip 2–3 mm in the liquid. The liquid is aspi-
rated by pressing the pipetting key. The arrow in the
display points upwards (aspiration).
Leave the tip immersed in the liquid for a few seconds,
so that the set volume is aspirated completely. This is
especially important when pipetting viscous media and
when using pipettes with large volumes.
Volume range
0.5 – 100 µl
100 – 300 µl
> 1000 µl
Immersion depth
Wait time
2 – 3 mm
1 s
2 – 4 mm
1 s
3 – 6 mm
3 s
Operating manual
4 Commissioning
53

Publicité

Table des Matières
loading

Table des Matières