Quality Control/ Internal Procedural Control; Results - Biohit ColonView Hb/Hp Test Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 46
(+20°C to +30°C) at least 10 minutes before testing.
2. Take the required number of test cassettes from the foil
packaging only immediately before performing the test. Mark
the cassette with the name of the patient or with another form
of identification.
3. Carefully shake the Sample Collection Tube to ensure that the
stool sample mixes properly with the saline solution.
4. Unscrew the protective white cap. Take a paper towel and
break the seal of the Sample Collection Tube with rotary
motion. Hold the collection container upright and slowly add 3
drops of the solution into the round sample window (S) of the
Test Cassette.
5. Read the results after 15 minutes. Strongly positive results
may be evaluated even sooner. Evaluate the result a maxi-
mum of 25 minutes after adding the solution into the round
sample window.

9.2 Quality Control/ Internal Procedural Control

The test contains a procedural control. A colored line
that appears in the control region (C) shows that each
test is performed correctly. It is easy for the background
to become slightly yellowish in color during testing, de-
pending on the color of the stool sample. This is ac-
ceptable, as long as evaluation of the test results is not
adversely affected. The test is invalid if the result lines
are not visible.

10. RESULTS

Test result is evaluated as "positive" if two lines appear.
Positive: 2 red/pink-colored lines appear; one in the control region
11

Publicité

Table des Matières
loading