For Crest Factor of 2.0 to 2.5, add 2.5%
to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0%
to accuracy.
CF≦ 2 600V//600A
(2) Resistance & Continuity
Function
Range
Ω
±(1% + 3 dgt)
400.0Ω
Overload protection : 600 Vrms
Max. open circuit voltage : 3V
Continuity check : Internal sounds acti-
vates if the resistance of the circuit under
test is less than 30Ω approximately.
(3) AC Current
Function
Range
±(1.9% + 7 dgt) *1
0~60.0A
A
60.1~400.0A
(50~60Hz)
±(1.9% + 5 dgt)
401~600A
0~400.0A
A
±(2.5% + 7 dgt) *1
(61~400Hz)
401~600A
Overload protection : 600 Arms
* 1 AC Conversion Type and additional
accuracy is same as AC Voltage.
Position Error : ±1% of reading.
26
Accuracy
Accuracy