Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
1-3 Electrical Specifications
Accuracy is ±(% reading + number of
digits) at 23°C ± 5°C < 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C, < 18°C,
> 28°C .
(1) Voltage
Function
Range
±(1.0%+ 5 dgt)
0~400.0Vrms
V
50Hz ~ 500Hz
400~600Vrms
*1
0~400.0 V
±(0.7% + 2 dgt)
V
400~600 V
Overload protection : 600 Vrms
Input impedance :
1MΩ // less than 100pF.
AC Conversion Type :
3350 : AC Conversion are average sens-
ing rms indication calibrated to the rms
value of a sine wave input.
3355 :
*1 : Reading less than 15% of full scale
(4000 digits) is needed to add 4 digits and
the settling time will increase.
AC Conversions are ac-acoupled, true
rms responding, calibrated to the rms
value of a sine wave input. Accuracies
are given for sine wave at full scale and
non-sine wave below half scale. For non-
sine wave add the following Crest Factor
corrections :
For Crest Factor of 1.4 to 2.0, add 1.0%
to accuracy.
25
Accuracy