Specifications
Model
Measurement method
Measurement range
Display accuracy
Measurement inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating conditions
Permissible storage conditions
Power supply
Battery life
Accessories
Classification
28
Guardy
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
Pulse 30 – 180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
4 x 30 memory spaces
L 155 mm x W 110 mm x H 70 mm
Approx. 395 g (without batteries)
22 to 42 cm
+5 °C to +40 °C, < 85 % relative air
humidity (non-condensing)
- 10 °C to + 55 °C, < 95 % relative air hu-
midity, 800 –1060 hPa ambient pressure
4 x 1,5 V
For minimum 900 measurements,
depending on the blood pressure level
and/or pump pressure
Instruction for use, 4 x 1.5 V AA alkaline
batteries., storage pouch
Internal supply, IP22, no AP or APG,
continuous operation, type BF applied part
AA alkaline batteries.