Measuring capacitance
IMPORTANT: If capacitors are
poled, the display unit handle
must be applied to the capaci-
tor's positive pole.
NOTE: No bar graph display is
given for capacitance measure-
ment.
Capacitance will not be meas-
ured, if the capacitor being
measured is charged. The
capacitor may, for example, be
discharged by pressing the
"LOAD" button.
Display hold (HOLD)
In awkwardly accessible places,
adverse lighting conditions. The
e.g. in switch cabinet, the dis-
"HOLD" function "freezes" the
play is often not clearly legible.
display to enable the user to
To overcome this problem, the
read and record the measured
display can be frozen during
values correctly after completing
voltage measurement with the
a measurement.
"HOLD" button
. Also, the dis-
play is illuminated to ensure that
it can be read easily even in
Selectable test probe light
The test probe light can be
turned on manually by pressing
the "ON/MODE" button. The
light only remains on while the
button is pressed.
Display lighting
The backlighting for the display
Under certain conditions, the
is automatically activated when
backgroundlighting may not be
the voltage tester is in operation.
visible in bright daylight.
LOAD button
Discharging capacitors
Attenuating reaction voltages
Pressing the "LOAD" button pro-
When measuring in AC voltage
vides a safe means of discharg-
networks, the reactive voltage
ing capacitors through the 2.5 k
component can be attenuated
ohm load resistor. Voltage will
by pressing the "LOAD" button.
be seen to fall on the display.