7 Guideline value for power settings

When performing the power settings, the user should generally
take into account the cutting speed, geometry of the electrode
Monopolar cutting:
Smooth cuts with thin needle electrode
Cutting with knife electrode, lancet or wire
loop
Monopolar coagulation:
Coagulation of large surfaces
Coagulation of small surfaces
Bipolar application:
Coagulation
Cut
Cut in water environment (TUR)
!
The recommended settings serve as guideline values only and
and properties of the tissue.
must be adjusted to the relevant situation.
In the event of unfavourable conditions, these values may
Recommended settings
5 - 40 W
50 - 175 W
Recommended settings
60 W
40 W
Recommended settings
30-60 W
40 - 60 W
50 - 75 W
deviate!