Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Voltage interruptions
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601
test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% U
T
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
T
AND 70% U
Single phase : at 0°
0% U
: 250/300 cycles
T
30 A/m
10 V/m
80 MHz to 2,7 GHz
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at 28 V/m
2450 MHz at 28 V/m
5240, 5500, 5785 MHz at 9 V/m
0,15 MHz – 80 MHz
6V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% MA at 1kHz
; 0,5 cycle
; 1 cycle
; 25/30 cycles
T
3V
54
Compliance level
± 8 kV
± 15 kV
± 2 kV
± 1 kV
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0% U
; 0,5 cycle
T
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% U
; 1 cycle
T
AND 70% U
; 25/30 cycles
T
Single phase : at 0°
0% U
: 250/300 cycles
T
30 A/m
10 V/m
80 MHz to 2,7 GHz
385 MHz at 27 V/m
450 MHz at 28 V/m
710, 745, 780 MHz at 9 V/m
810, 870, 930 MHz at 28 V/m
1720, 1845, 1970 MHz at 28 V/m
2450 MHz at 28 V/m
5240, 5500, 5785 MHz at 9 V/m
3V
0,15 MHz – 80 MHz
6V in ISM bands and amateur radio bands
between 0.15 MHz and 80 MHz
80% MA at 1kHz