The value of the overpressure reached can be read on the manometer.
Adjust the overpressure between 10 and 50 mbar (0.15 and 0.75 PSI)
according to the sensations of the patient and the indications of the
doctor.
If the needle on the gauge does not move, check the airtightness of the
nasal plug on the nose.
Start again the operation according to the doctor's indications.
Optimal duration: 6 min., that is to say 15 effective swallowings.
d) After each use
At the end of the treatment:
Stop the device (switch on O).
Disconnect the electric cord from the mains.
Disconnect the tubes (pressure and vibrations) from the compressor.
Follow the cleaning and disinfection instructions.
5. AEROSOLTHERAPY
a) Performances of the MANOSONIC nebulizing system
See granulometric curve opposite
MMAD: 4.5 µm
Cumulative particle size distribution of the 2.5% NaF
aerosol produced by the MANOSONIC system
Information regarding filling volumes, inhalable mass and air flow are mentioned on the nebulizing kit
packaging.
6. DEVICE CLEANING
a) Device cleaning
After each use, cleaning is limited to the external parts with a non-abrasive sponge soaked with alcohol or
with a disinfecting product recommended by the manufacturer (Anios Surfa'Safe Premium).
18
5
Aerodynamic diameter (µm)