9. Specifications
Output signal 4 ... 20 mA et HART
Transmitter model (selectable versions)
Data sheet
EN
Output
4 ... 20 mA
HART
protocol
®
Galvanic isolation
Measuring insert (replaceable)
Material
Diameter
Spring travel
Response time (in
water, per EN 60751)
Neck tube
Material
Connection thread to
the thermowell
Connection thread to
the head
Neck length
14) Protect the temperature transmitter from temperatures > 85 °C.
26
protocol
®
T16
TE 16.01
x
-
x
Ni alloy 2.4816 (Inconel 600), others on request
Standard: 3 mm, 4.5 mm, 6 mm, 8 mm
Option (on request): 1/8" (3.17 mm), 1/4" (6.35 mm), 3/8" (9.53 mm)
approx. 20 mm
t
< 5 s
t
< 10 s (measuring insert diameter 6 mm: The thermowell
50
90
required for operation increases the response time dependent upon
the actual parameters for the thermowell and the process.)
Stainless steel 1.4571, 316, 316L
G 1/2 B, G 3/4 B, 1/2 NPT, 3/4 NPT, M14 x 1.5, M18 x 1.5, M20 x 1.5,
M27 x 2
M20 x 1.5 with counter nut
M24 x 1.5 with counter nut
150 mm, standard neck length
200 mm
250 mm
other neck lengths on request
WIKA operating instructions models TR12, TC12
14)
T32
TE 32.04
x
x
x
1/2 NPT
3/4 NPT
TIF50, TIF52
TE 62.01
x
x
x