SPA-22413 Manual BC_DTP.qxd
Problem
The water splash is
making a mess.
The water pressure
from the eartip
seems too high or
too low.
3/7/02
9:36 AM
Possible Cause
Excess water pressure
from the water line.
Installation of eartip has
caused a problem.
There is leakage from the
chamber aerator and/or
adapter.
The return water exit port
is clogged on the bottom
of the chamber.
There is an internal
malfunction.
13
Page 13
Solution
Ensure that the chamber
aerator is attached to the
bottom of the chamber.
Apply a new eartip with the
tab side pointing upwards.
Tighten the aerator and/or
adapter with a pair of pliers.
Remove the chamber from the
faucet and dislodge any
foreign materials from the exit
port. Proceed with the normal
cleaning procedure.
With an eartip attached,
spray water from the handle
into a graduated cylinder or
measuring cup for exactly 20
seconds. If there was less than
90 mL or more than 110 mL of
measured water, contact your
Schering-Plough Animal Health
Office (see page 97).