Troubleshooting
If your SPAH Auriflush System is experiencing problems, consult the following table to
find the symptom, the likely cause and the solution. If the information in this table does
not describe or solve the problem you are having, contact your Schering-Plough Animal
Health Office (see page 97).
Problem
No water is coming
out of the handle
even though I am
pulling the actuator.
Excessive water
leaking from the
ear.
Lack of a clean
water stream
coming from eartip.
3/7/02
9:36 AM
Possible Cause
The temperature shut-off
safety feature has acti-
vated due to the hot water
temperature.
There is a kink in the
tubing.
The tubing on the INSIDE
of the handle is restricting
water flow.
The water is not turned on
high enough.
Eartip attachment
problem.
Tubing connection
problem.
Incorrect technique.
Installation of eartip has
caused a problem.
12
Page 12
Solution
Run COLD water through the
system for 10-30 seconds. This
will reset the temperature shut
off safety feature and allow
water to flow again.
Make sure there are no bends
in the tubing restricting the
flow of water.
Prime the actuator 25 - 50
times until water begins to
flow normally
Turn the cold water on full in
order for the suction system to
work properly (but still within
the right temperature range).
Replace the eartip with the tab
side pointing upwards.
Ensure that both tube fittings
are fit snugly into the chamber.
Check the Quick Reference
Guide to verify you are aiming
the water correctly.
Apply a new eartip with the
tab side pointing upwards.