Warranty Information - VWR Mini Centrifuge Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 9
Troubleshooting
Error
Unit does not operate
Excessive Noise or
Vibration
Lid will not open when
button is pressed
Unit stops running
without pressing front
button

Warranty Information

VWR Mini Centrifuge 2 Year Warranty
VWR Mini Centrifuge is warranted to be free of defects in materials and workmanship
for 24 months from the date of purchase, providing that:
• The unit has not been opened or attempted to be opened by the end user.
• The user ensures that it has been carefully cleaned and decontaminated before returning.
• The unit has been operated according to the instructions and not in any improper way.
• The unit has not been damaged due to negligence, accident or used under abnormal
conditions (temperature, corrosives, unbalanced rotors, lack of cleaning, liquid
introduced into housing).
• Normal wear and tear of unit.
In order to qualify for the warranty repair; the customer must return the warranty card
supplied with the unit that includes all of the information required to validate proof of
purchase. Should warranty service be required, repair or replacement will be provided by
VWR without charge. Freight to ship the unit to VWR will be the responsibility of the
purchaser. There are no other warranties or guarantees, implied or otherwise expressed
including any warranties of application or merchantability.
In no event will this warranty obligation exceed price of product. This includes liability
claims of loss of use, lost profits or any exemplary or punitive damages.
Cause
Loose Cord.
On / Off Switch.
Lid not closed.
Blown Fuse.
Rotor not balanced.
Rotor not installed
properly.
Foreign object in unit.
Power is off.
Loose Cord.
Lid not completely
closed.
60 minute time limit
expired.
5
VWR Mini Centrifuge — Operator's Manual
Solution
Check Cord.
Turn switch to "On" position.
Close lid until it clicks and
locks in place.
Replace Fuse.
Check tube fluid levels and
symmetry.
Remove & reinstall rotor per
above instructions.
Check centrifuge bowl for
loose tubes or objects.
See Above.
Open lid manually.
Check Cord.
Close lid until both pins are
locked.
Centrifugation cycle is
complete.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

10067-58810081-04010081-04210081-044

Table des Matières