19
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Eolo aerosol can be used in the following electromagnetic environment. The Customer and/or
the user of the Eolo aerosol shall make sure that the appliance is used in such environment.
Immunity test
Electrostatic
discharge (ESD) -
EN 61000-4-2
Fast transient / burst
EN 61000-4-4
Surge EN 61000-4-5
Voltage dips, short
outages and
voltage variations
EN 61000-4-11
Network frequency
magnetic field
EN 61000-4-8
Note: U T is the value of the supply voltage
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Eolo aerosol can be used in the following electromagnetic environment. The Customer and/or
the user of the Eolo aerosol shall make sure that the appliance is used in such environment.
Immunity
Level set out by
test
EN 60601-1-2
Conducted
3Vrms 150kHz
immunities
to 80MHz
EN 61000-4-6
(for non-life-
supporting appliances)
Radiated
10 V/m 80MHz to
immunity
2.7GHz
EN 61000-4-3
(for non-life-equipment
appliances)
Standard of proof
± 8kV contact
± 15kV air
± 2kV power supply
± 1kV signal cables
± 0.5kV e± 1kV
differential mode
5% U T (>95% dip
in U T ) per 0.5 cycle
40% U T (60% dip
in U T ) per 5 cycles
70% U T (30% dip
in U T ) per 25 cycles
<5% U T (>95% dip
in U T ) per 5 s
30 A/m
Level of
conformity
V
3 V rms
E
10 V / m
ENGLISH
Level of
conformity
The appliance
does not alter
its status
The appliance
does not alter
its status
The appliance
does not alter
its status
--
The appliance
does not alter
its status
Electromagnetic Environment - Guide
=
Portable and mobile RF communication
1
devices should be used no closer than the
separation distance from any part of the
Eolo device, including cables, calculated
from the equation applicable to the fre-
quency of the transmitter.
=
1
Recommended separation distances
d= [3.5 / V
d= [12 / E
d= [23 / E
P is the maximum rated output power of
the transmitter in Watt (W) according to
the manufacturer of the transmitter and d
is the recommended separation distance
calculated in meters (m).
Guide to the electromagnetic
environment
Floors should be in wood, cement
or ceramics. If floors are covered
by synthetic material, the relative
humidity should be at least 30%
The power supply should be that
of a typical commercial premise
or hospital.
The power supply should be that
of a typical commercial premise
or hospital.
The power supply should be that
of a typical commercial premise
or hospital.
If the user of the Eolo aerosol re-
quires the appliance to work con-
tinuously, it is recommended to
use it with an uninterruptible pow-
er supply.
The power supply should be that
of a typical commercial premise
or hospital.
] √P
1
] √P 80MHz to 800MHz
1
] √P 800MHz to 2,7GHz
1