8. Rectifying faults

Error messages can occur when
• the blood-pressure value is unusually high or low („EE"
appears in the display),
• you move or talk during the measurement („EE" appears
in the display),
• the cuff tube is not properly inserted („E1" appears in the
display),
• inflation takes longer than 20 seconds („E1" appears in
the display),
• the inflation pressure is higher than 300 mmHg („E2"
appears in the display),
• an error occurs when storing the measured values („E3"
appears in the display),
• Exceeding measurement range („Er" appears in the dis-
play).
In the about cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk. Put the batteries back in if neces-
sary, or else replace them.
9. specifications
Type:
BC 16
Measuring method:
non-invasive, oscillometric blood
pressure measurement at the wrist,
for wrist circumferences of
13.5 – 22 cm
Display range:
Measuring range:
Accuracy of the
pressure reading:
Measuring uncertainty: (max. permissible standard devia-
Power supply:
Classification:
Ambient conditions:
Storage/transport:
Key to symbols:
These specification are subject to change without notice for
purpose of improvement.
17
0 – 299 mmHg
systolic: 30 – 260 mmHg/
diastolic: 30 – 260 mmHg/
pulse: 40 – 199 beats/min
systolic ±3 mmHg/
diastolic: ±3 mmHg/
pulse: ±5 % of the displayed value
tion according to clinical testing:
systolic: 8 mmHg/
diastolic: 8 mmHg
2 x 1.5 V Micro
(Alkaline Type LR 03)
Application part Type BF
+10 °C – +40 °C max. 40 – 85 %
atmospheric humidity
-10 °C – +60 °C max. 10 – 90 %
atmospheric humidity
Application part Type BF
Important! Read the operating
instructions