Principle Of Meas., Wiring Assignment: Wa10 - HBM WA 2 Notice De Montage

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12
É É É É É É
-
Fig. 3.4: General electrical wiring‐diagram, half bridge with 40mV/V
For information on using other transducer connection types, refer to the
operating manual for the chosen amplifier (connection diagram WA, see types
of connection).
Fig. 3.5: Lemo connector pin assignment (solder side of male cable connector)
3.5 Principle of meas., wiring assignment: WA10...WA500
0
É É É É
Fig. 3.6: General electrical wiring‐diagram, full bridge
HBM
+
WH (white)
BU (blue)
GN (green)
BK (black)
GY (gray)*
Cable shield
WH
BU
GN
RD
80mV/V, 6‐pin
)
WH (white)
BU (blue)
GN (green)
RD (red)
BK (black)
GY (gray)
Cable shield
Measurement signal ())
Excitation voltage (+)
Sensor circuit (+)
Excitation voltage (-)
Sensor circuit (-)
Transducer housing
* for the high and low temperature version:
violet
BK
GY
Measurement signal ())
Excitation voltage (+)
Sensor circuit (+)
Measurement signal (*)
Excitation voltage (-)
Sensor circuit (-)
Transducer housing