guiDancE & manufacturEr'S
DEclaration
Guidance and Manufacturer's Declaration – Emissions All Equipment and
Systems
This device is intended for use in the electromagnetic environment specified below. The
customer or user of this device should assure that it is used in such an environment.
IEC 60601
Immunity Test
Test Level
Power
Frequency
50/60Hz
3A/m
Magnetic Field
IEC 61000-4-8
3 Vrms
Conducted RF
from
IEC 61000-4-6
150 kHz to
80 MHz
3 V/m
Radiated RF
80 MHz to
IEC 61000-4-3
2.5 GHz
For transmitters rated at a maximum output power not listed above, the recommended separation
distance D in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Compliance
Level
Electromagnetic Environment - Guidance
Power frequency magnetic fields should be
3A/m
that of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications
equipment should be separated from the
device by no less than the recommended
V1 = 3
separation distances calculated/listed
Vrms
below:
D=(3.5/V1)
D=(3.5/E1)
D=(7/E1)
Where P is the maximum power rating in
watts and D is the recommended
separation distance in meters.
E1 = 3V/m
Field strengths from fixed transmitters, as
determined by an electromagnetic site
survey, should be less than the compliance
levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.
√
P
80 to 800 MHz
√
P
√
800 MHz to 2.5 GHz
P
SE-DV54-1
EN
33