S O T O X A™ T E S T K I T I N S T R U C T I O N S F O R U S E
O P E R AT I N G C O N D I T I O N S
The operational temperature for running a test is between 5°C to 35°C.
CUTOFF LEVELS
Drug Class
Cannabis (THC)
I N T E R P R E TAT I O N O F R E S U LT S
Interpretation is carried out by the analyzer, which will display either a negative, positive or invalid result for the drugs which
have been tested.
Negative:
This indicates that the drug concentration in the oral fluid sample is below the designated cutoff level for that
specific drug.
Positive:
This indicates that the drug concentration in the oral fluid sample exceeds the designated cutoff level for that
specific drug.
Invalid:
A result could not be provided for that specific drug.
L I M I TAT I O N S
1. This assay is for forensic use only and provides a preliminary qualitative analytical test result.
2. Confirm positive results using a more specific alternative chemical method; GC-MS or LC-MS is preferred.
3. It is possible that technical or procedural errors, as well as other interfering substances in the oral fluid sample may cause
erroneous results.
4. This assay is for the analysis of oral fluid samples only collected by the SoToxa Oral Fluid Collection Device. Testing with
other sample types may not give valid results.
5. A positive result indicates presence of the drug or its metabolites but does not indicate level of intoxication, administration
route or concentration in the oral fluid.
6. A negative result may not necessarily indicate drug-free oral fluid. Negative results can be obtained when a drug is present
but below the cutoff level of the test.
7. The test does not distinguish between illegal and prescription use.
T E C H N I C A L S U P P O R T
For product technical support and advice, please refer to: abbott.com/roadside
6 | EN
Abbreviation
THC
Cutoff (ng/mL)
25