S O T O X A™ T E S T K I T I N S T R U C T I O N S F O R U S E
S O T O X A T E S T P R O C E D U R E
A
S E T U P T H E C A R T R I D G E
1
Press the power button to switch the analzyer on. Wait for the initializing process to complete.
2
When prompted enter the operator or supervisor passcode, followed by the operator name/ID.
3
Remove a new test cartridge from its foil wrapper.
4
Check that the silica gel pack is present, intact and is yellow.
If the silica gel is missing, open or is green, discard the test and start again with a new cartridge.
Report the problem to Technical Support.
5
When prompted by the analyzer screen, insert the test cartridge.
The analyzer will check that the cartridge is valid.
Make sure to keep the analyzer horizontal and still at all times.
B
C O L L E C T T H E D O N O R S A M P L E
6
Unwrap a new Collection Device from the packaging.
Ensure that they hold the collection device by the plastic stem, and then place it in the mouth.
7
Actively swab the collection device around gums, tongue and inside the cheek.
8
Continue swabbing until the sample presence indicator starts to turn blue.
C
R U N T H E T E S T
9
Insert the collection device into the test cartridge (in the analyzer).
Gently push all the way into the cartridge to the stop position.
10
The analyzer will now test the sample. The test time will be displayed on the screen.
Make sure to keep the analyzer horizontal and still at all times.
11
The results will be displayed on the analyzer screen. If it has been enabled, a donor questionnaire
will begin after pressing 'OK'. If required, the results can be printed.
12
Please ensure that the printer is connected to the analyzer and has been switched on before printing.
To skip this step press 'NO'.
13
The test cartridge and collection device can now be removed from the analyzer.*
Remove by holding the sides of the test cartridge and pull gently. Do not attempt to pull the collection device
or remove from the test cartridge.
* Take care when removing the test cartridge as the underside of the cartridge may be hot.
Handle the test cartridge by the side edges.
4 | EN