5 Operation
5.1.3 Evaluating a test result with control
solution
The permissible range for the control solution reading
is indicated on the label of the test strip container. Your
test result must lie within the range indicated. Make
sure that you compare the test result with the correct
range. If the control test result lies within the range
indicated on the test strip container, then the test strips,
the GlucoDock module and your mobile device are
working accurately.
NOTE
If you save test results with the control
solution, they are included in your
regular measuring data! To prevent
this, select control solution mode.
If the control test result does not lie within the range
indicated on the test strip container, then the following
options will be displayed to rectify the problem:
26
GB
10. After comparing the test
value with the value specified
on the test strip container,
remove the test strip and
dispose of it.