The P
LED handpiece is equipped with a
IEZOTOME
piezoelectric transducer.
It has a light function for optimal lighting of the
procedure site.
This light function is achieved by a LED ring consisting
of high-performance LEDs, giving a considerably
longer life than conventional optical fiber or bulb
technologies.
In addition to the performance of its piezoelectric
transducer and its optical properties, the front part
of the P
LED handpiece can be completely
IEZOTOME
disassembled for optimal cleaning and sterilizing.
I - INSTALLATION
1. 1 PIEZOTOME LED HANDPIECE-CORD ASSEMBLY
CONNECTION
(Please refer to P
IEZOTOME
- Align the mark on the cord connector with the mark
on the left-hand or right-hand socket of the unit.
- Insert the connector in the socket until you hear the
"click".
1. 2 TIPS MOUNTING
The P
LED handpiece is a high technology
IEZOTOME
instrument requiring precautions for its use. The tips
must be screwed and moderately tightened (¼ turn)
by means of the P
IEZOTOME
universal wrench of Satelec (Fig. 1).
CAUTION: Correct tip vibration requires perfect
tightening without overtorquing beyond the thrust
point. Remove the tip after each procedure to avoid
self-jamming.
unit user's manual)
specific wrench or the
II - MAINTENANCE
Precaution: rinse the water circuit of the P
LED handpiece-insert assembly under running water
for 20 to 30 seconds after each use.
Caution: do not connect the P
cord assembly from the unit when the instrument is
powered and the control pedal is activated.
2. 1 MAINTENANCE OF THE PIEZOTOME LED
HANDPIECE-CORD ASSEMBLY (Fig. 2)
a) Pre-disinfection/cleaning of the P
handpiece-cord assembly
The P
LED handpiece-cord assembly must
IEZOTOME
not be cleaned in an ultrasonic bath.
Disconnect the P
IEZOTOME
from the unit and remove the insert.
Remove the tip at the front of the P
(Fig. 2-A).
The front part of the P
unscrewed, for easier access to the threading and
efficient and easy cleaning, using a cotton swab
soaked with alcohol.
Remove the optical guide from inside the P
LED handpiece nose (Fig. 2-A).
Disconnect the light ring from the P
handpiece, pulling it gently (Fig. 2-A).
Clean and disinfect the body and cord, using cleaning
/ disinfecting towelettes with alcohol, amphoteric
and biguanide base (S
use (refer to the manufacturer's instructions), for at
least two minutes, paying particular attention to the
screw thread cavity; the nose; and the optical guide
and LED ring.
Allow the product to act for at least 15 minutes.
Use towelettes that bear the EC marking, or that are
in compliance with any standard required under the
applicable national regulations.
4
LED handpiece-
IEZOTOME
IEZOTOME
LED handpiece-cord assembly
IEZOTOME
LED handpiece can be
IEZOTOME
IEZOTOME
™ type towelette) ready for
EPTOL
IEZOTOME
LED
handpiece
IEZOTOME
LED