8.0 Operator Control Panel

1. ON/STANDBY: Switches Control Unit between On and STANDBY modes.
2. POWER INDICATOR LED: Indicates AC power is present to the Control Unit.
3. SOFT (-): Decreases the inflation level of the bed while Control Unit is in manual mode only. Pressing
once will decrease the setting by one unit. Pressing and holding will rapidly change the setting.
4. COMFORT CONTROL INDICATOR: Displays the current inflation level.
5. FIRM (+): Increases inflation level of the bed while in manual mode only. Pressing once will increase
the setting by one unit. Pressing and holding will rapidly change the setting. Button is inactive in
Automatic mode.
6. MAX INFLATE: Selects/deselects the max inflate mode (maximum mattress firmness). The indicator
within the button will flash while max inflate mode is active. After 30 minutes, max inflate mode will
cancel. At power-up, max inflate is active for 30 seconds.
7. LOCKOUT: activates or cancels lockout mode. Press and hold for three seconds to activate or cancel
mode. The indicator within the button will light while lockout is active. In lockout mode, only the MAX
INFLATE button is operational. All other buttons are temporarily inactive.
8. AUTO: selects/deselects the automatic mode of operation. The indicator within the button will light
while automatic mode is active.
Automatic mode will not activate if the M2500S / Auto Aire Select (with sensor cell option) mattress
sensor cord is not connected to the jack on the right side of the Control Unit.
Note: Connecting the mattress sensor cord while the unit is not in automatic mode will cause the Control
Unit to enter automatic mode by itself.
The AUTO button can be used to exit automatic mode at any time.
9. AP: selects the alternating low pressure mode of operation. The indicator within the button will light
while AP mode is active. Note: AP mode is temporarily suspended in max inflate mode.
10. ALARM SILENCE: Temporarily silences the audible alarm for 15 minutes.
11. POWER FAIL INDICATOR: Flashes when power supplied from line cord is interrupted while the Control
Unit is operating. To avoid an inadvertent power fail alarm, place the unit in STANDBY mode before
disconnecting power from Control Unit.