Weight
Approx. 115 g (without batteries)
Cuff size
135 to 195 mm
Permissible operating
+10 °C to +40 °C, ≤ 85 % relative air
conditions
humidity (non-condensing)
Permissible storage
-20 °C to +70 °C, ≤ 85 % relative air
conditions
humidity, 700 –1060 hPa ambient pres-
sure
Power supply
2 x 1.5 V
Battery life
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Accessories
Instruction for use, 2 x 1.5 V AAA batter-
ies, storage box
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus-
tomer Service address or found at the end of the instructions
for use.
• This device is in line with the EU Medical Devices Direc-
tive 93/42/EC, the „Medizinproduktegesetz" (German Medi-
cal Devices Act) and the standards EN 1060-1 (non-inva-
AAA batteries
18
sive sphygmomanometers, Part 1: General requirements),
EN 1060-3 (non-invasive sphygmomanometers, Part 3: Sup-
plementary requirements for electro-mechanical blood pres-
sure measuring systems) and IEC 80601-2-30 (Medical elec-
trical equipment – Part 2 – 30: Particular requirements for the
safety and essential performance of automated non-invasive
blood pressure monitors).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful life.
If using the device for commercial medical purposes, it must
be regularly tested for accuracy by appropriate means. Pre-
cise instructions for checking accuracy may be requested
from the service address.