Immunity test
Test / Compliance level
< 0% U
Voltage dips, short
45° steps
interruptions and
voltage variations
0% U
on power supply
70% U
input lines
< 5% U
IEC 61000-4-11
cycles
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
3 V
Conducted RF
150 kHz to 80 Mhz
IEC 61000-4-6
6 V
in ISM & amateur radio
bands
28
Electromagnetic environment – guidance
for 0,5 cycle at
T
Mains power quality should be that of a typical commercial or hospital
environment. If the user of this product requires continued operation during
for 1 cycles
power mains interruptions, it is recommended that the product is powered
T
from an un-interruptible power supply or a battery.
for 25 / 30 cycles
T
for 250 / 300
T
U
is the a. c. mains voltage prior to application of the test level.
T
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which this product is used exceeds
the applicable RF compliance level above, this product should be observed
to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating this product.
Interference may occur in the vicinity of equipment marked with the following