Preparation For Use - GCE EASE II Mode D'emploi

Masquer les pouces Voir aussi pour EASE II:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
In case of storage at a temperature below -20 °C or above 60°C do not operate the demand
valve until it has been allowed to change its temperature inside operating temperature limits.
For the demad valve intended to be used with mixture of gases O2+N2O, the lowest operat-
ing temperature is +5°C.
Do not use the device without being properly trained!
EN
All gases under pressure can be hazardous and should be treated with caution.
Do NOT use oil or grease (unless specifically approved) on any part of an EASE II system.
Ensure hands are clean and free from oil and grease.
Keep system away from sources of ignition at all times.
Do NOT tamper with the equipment.
Do NOT try to connect an EASE II valve to terminal unit/quick connector of the incorrect gas
type.
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider must ensure that all
personnel handling the product are provided with the operating instructions & performance
data.
Do not use the product without properly familiarization of the product and its safe operation
as defined in this Instruction for use. Ensure user is aware of particular information and knowl-
edge required for the gas in use.

5. PREPARATION FOR USE

5.1. CONNECTION OF THE HOSE TO THE GAS SOURCE
Connect the supply
to the source of gas. For
more information regarding
gas - specific medical quick
connectors, see to Appendix
2.
Regularly inspect the hose
during use to ensure that
the hose is not torn, bent,
twisted,
excessive pull.
Regularly check the supply
of gas or the contents in the
cylinder.
5.2. FITTING
SCAVENGING ADAPTOR
When using the scavenging adaptor, fit it to the demand valve by pushing it over the patient
port. (see Fig. 1).
Then connect the exhalation pipe to scavenging adaptor.
NOTE: A scavenging adaptor should be fitted when EASE II for O2/N2O is used in enclosed
spaces. Failure to fit a scavenging adapter can result in supervisory personnel becoming
drowsy.
hose
or
affected
by
Mask
Antibacterial
filter
Fig. 1: Fitting scavenging adaptor, filter and mask
4/88
Scavenging
adaptor
Test button