ENVIRONMENTAL REqUIREMENS

As a precision instrument, the analytical balance requires an environment
which is free from excessive air flow, corrosive, vibration and temperature
or humidity extremes.The mentioned factors will affect displayed weighing
readings.
Keep weighing ambient clean and dry all time;
The best operating temperature is about 20ºC / 68ºF at about 50%
relative humidity.
Use a stable AC main.
Don´t place the balance:
– In direct sunshine.
– Next to open windows or a door causing draft or rapid temperature
charges.
– Near heater or air conditioner.
– Near vibrating, rotating or reciprocating equipment.
– Near magnetic field or equipment which generates magnetic field.
– On an unstable weighing table.
– In areas where the danger of explosion exists.
0,2 mg
About 5 sec.
2 ppm
External
20 - 25 ºC
90 mm
320 x 300 x 470 mm
AC 110V-230V 50Hz-60Hz
Automatic internal
7