Fuse Replacement; Troubleshooting - Hach Lachat BD40 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20

Fuse replacement

Fuses are not user-serviceable items. The need for fuse replacement in controllers indicates severe
technical failure and is therefore considered to be a service activity. If a blown fuse is suspected,
contact Technical Support.

Troubleshooting

Problems with the digestion block cause an alarm to sound and a safety message to show on the
controller display. The controller will no longer apply heat to the digestion block and it will start to
cool. Refer to
Table
5.
After the problem is resolved:
• Push CONTINUE to clear the message and let it occur again if necessary.
• Push SUSPEND to disable the message until the system is powered off.
Message
Overtemp. Protection–
Check Block Temp.
18 English
Table 5 Safety messages
Possible cause
The block is too hot when the sample
is added.
The voltage switch on the back of the
block is on the wrong voltage setting.
An exothermic reaction has occurred.
Solution
Let the block cool to room temperature
before the next digestion starts.
Change the voltage switch to the correct
setting.
Make sure that the method in the controller
matches the written method. Repeat the
test.

Publicité

Table des Matières
loading

Table des Matières