FR
LE DISPOSITIF EST DESTINÉ EXCLUSIVEMENT AU LAVAGE ET/OU À LA
DÉSINFECTION THERMIQUE DES INSTRUMENTS CHIRURGICAUX
RÉUTILISABLES EN MESURE DE RÉSISTER À UNE TEMPÉRATURE D'AU
MOINS 80 °C OU BIEN 90 °C SELON LE CYCLE SÉLECTIONNÉ.
IL N'EST PAS DESTINÉ AU LAVAGE ET À LA THERMO-DÉSINFECTION DE
DISPOSITIFS MÉDICAUX THERMOLABILES.
IL N'EST PAS DESTINÉ AU LAVAGE D'INSTRUMENTS MÉDICAUX QUI
PRÉSENTENT DES CAVITÉS PROFONDES AVEC OUVERTURE RÉDUITE
(ÉTANT DONNÉ QUE LE NETTOYAGE ET LA DÉSINFECTION DES
PARTIES INTÉRIEURES DE LA CAVITÉ NE SONT PAS GARANTIS).
IL N'EST PAS DESTINÉ AU LAVAGE D'INSTRUMENTS ROTATIFS OU
ÉLECTRONIQUES, DÉFINIS APPROPRIÉS PAR LE FABRICANT DE
MANIÈRE EXPLICITE
IL N'EST PAS DESTINÉ AU LAVAGE D'INSTRUMENTS EN ALUMINIUM
NON PROTÉGÉ OU AVEC ANODISATION DÉCORATIVE.
CYCLES
PRÉLAVAGE
DÉSINFECTION
2 l
D90
LAVAGE W
2 l
PRÉLAVAGE
2 l
INSTRUCTIONS D'UTILISATION
ATTENTION
CONSOMMATION D'EAU PENDANT LES PHASES DU CYCLE
LAVAGE
6lt
6lt
/
RINÇAGES
THERMO-DÉSINFECT.
n 3 x 2 l
1,5 l
n 3 x 2 l
/
/
/
63