Troubleshooting

E
Symptom
Pump does not power
up when the stand-by
power switch is turned
on.
No fluid flowing
through tubing.
Fluid flow rate is not
consistent.
User Information Code
consistently displayed.
Table 1: Troubleshooting Guide
Possible Cause
♦ Power cord is unplugged.
♦ Suspect power cord.
♦ Blown fuse(s).
♦ Fluid bag is empty.
♦ Tubing line is severely
bent or kinked.
♦ Clamps not open.
♦ Fluid bag is empty.
♦ Tubing line is severely
bent or kinked.
♦ Clamps not open.
E-16
Corrective Action
♦ Plug power cord into the pump and/
or a properly grounded receptacle.
♦ Replace power cord.
♦ Check fuses. Replace.
♦ Replace fluid bag.
♦ Remove all severe bends and/or
kinks. If tubing is damaged, replace
tubing set.
♦ Open clamps.
♦ Replace fluid bag.
♦ Remove all severe bends and/or
kinks. If tubing is damaged, replace
tubing set.
♦ Open clamps.
♦ Ensure no buttons are being
depressed on the pump or remote.
Verify the cassette is seated
properly or completely removed.
Cycle power off, wait at least five
seconds, then cycle back on to clear
the code. If code is still present, or
returns, return pump to Linvatec for
service.