2.2

Pole Mount Installation

E
CAUTIONS:
1. The base of the infusion pole in use must
be at least 21.5 inches (54.5 cm) in
diameter.
2. Do not mount the pump higher than
49 inches (1.25 m) from the floor or
pump and pole may tip.
3. Do not move the infusion pole with saline
bag(s) attached. Remove bag(s) prior to
transporting.
4. When moving the pump, grasp the pole,
do not pull using the pump.
1. To install the pump onto the infusion pole,
support the pump in one hand and insert the
infusion pole into the integrated pole
mounting bracket on the back of the pump.
2. Position the pump at approximately the
same height as the inflow portal and
securely tighten the clamping knob.
3. The bag hanger should be no more than 32
inches (81.25 cm) above the pump with the
bag height not exceeding 76 inches (1.9 m)
from the floor.
NOTE:
When removing the pump from the
pole, always support the unit by placing your
hand underneath the pump prior to loosening
the clamping knob.