Procedure:
Switch off the balance.
Press adjustment switch and turn on weighing scale by operating foot switch
(chap. 3)..
Wait until the size of the adjustment weight (see chap. 1) required is displayed.
Place adjustment weight in the centre of the weighing platform. Wait until „Ld 0"
is shown.
Take away adjustment weight. Ensure that there are no other objects on the
weighing pan.
Wait for a few seconds until „PASS" is displayed.
After successful adjustment the balance automatically returns to weighing
mode.
18
MPD-BAPS-e-1811