6. Commissioning, operation
Set the display
Indication
Simultaneously press and hold buttons 1 and 3 for five seconds. The indicator
■
switches to "UNIT".
Using buttons 2 and 3, select the required unit.
■
Confirm the unit set, using button 1. The display reverts to "UNIT".
■
Briefly press button 1 in order to bring up the menu point "dP" (decimal point).
■
Using buttons 2 and 3, select the required position for the decimal point.
■
Confirm the position set using button 1. The display reverts to "dP".
■
Briefly press button 1 in order to bring up the menu point "di.Lo" (display low).
■
Using buttons 2 and 3, set the display value to be either 4 mA or 0.1 V
■
Confirm the value set using button 1. The display reverts to "di.Lo".
■
Briefly press button 1 in order to bring up the menu point "di.Hi" (display high).
■
Using buttons 2 and 3, set the display value to be either 20 mA or 10.1 V
■
Confirm the value set using button 1. The display reverts to "di.Hi".
■
By pressing button 1 briefly, the settings are saved. A successful 'Save' is signalled by
■
the display switching off briefly.
WIKA operating instructions attachable indicator, models A-AS-1, WUR-1
Unit
5 s
dP
di.Lo
di.Lo
Select unit
Confirm
Set the decimal point
Confirm
Set the smallest display
value
Confirm
Setting the largest
display value
Confirm
15
GB