5.3 Mounting and dismounting the AFI5 cable
5.4 Mounting the AFIx
WARNING
Danger to health from contaminated medium
Only use weld-in sleeves or adapters from Baumer.
f
Do not seal the process connections with Teflon tape (PTFE) or elastomer.
f
Welding work must only be carried out by welders trained in the area of hygiene.
f
9 Hole for mounting the sensor is easily accessible and dry.
9 Vessel is drained.
9 Mounting location and sensor orientation meet the conditions in chapter "5.1 Mounting conditions" on
page 5.
Check leak-tightness of the sleeve.
f
Check leak-tightness of the cable glands or M12 plug.
f
Ensure that the housing cover is fastened properly.
f
www.baumer.com
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Mounting the AFI5 cable
To plug the cable in, align the 2 red points.
f
A locking mechanism ensures that the cable
cannot be pulled out unintentionally.
Dismounting the AFI5 cable
Press both ends of the locking mechanism
f
together.
Pull the cable out.
f
Mount weld-in sleeve or adapter as follows:
f
3-A mark or arrow points upwards
Leakage hole points downwards
Hygienically and internally flush
Grind welding to Ra ≤ 0.8.
f
Screw in sensor.
f
Tightening torque: 20 ... 25 Nm
Operating Instructions
7 / 52