Inductive conductivity transmitter
Some weld-in sleeves (e.g. ZPW2-626) must be mounted in the correct angle to allow automatic draining.
Example of mounting with weld-in sleeve ZPW2-626
170°
5.2 Changing orientation of the display
Depending on the mounting location and orientation of the sensor, the facing and orientation of the display
can be adjusted accordingly.
0...345°
6 / 52
Conductivity measurement
1
Operating Instructions
190°
Open housing by unscrewing the cover.
f
Lift display from housing.
f
If desired, turn head of sensor:
f
Loosen the 2 screws (1) inside with a 2 mm
Allen key.
Turn head of sensor to the left (max. 345°).
Tighten the 2 screws (1) inside with a 2 mm
Allen key.
Make sure not to damage connecting ribbon
f
cable and put the display back in housing in the
desired orientation.
Close housing by screwing on the cover.
f
www.baumer.com